Use the Electrode Surface node to model an electrochemical electrode-electrolyte interface between an electrolyte domain and an electrode boundary where the electrode is not included explicitly as a domain in the model geometry. Set the electric potential of the electrode or specify a current condition that the potential of the electrode shall fulfill, and use subnodes to specify the 
Electrode Reaction and the 
Double Layer Capacitance at the interface.
 
    Use the Add (

) and 
Delete (

) buttons as needed in the table to control the number of species.
 
    The Density and 
Molar mass, in conjunction with the reaction rates and stoichiometry, defined in the 
Electrode Reaction subnodes, determine the normal electrode growth rate. 
 
    When the Solve for species concentrations variables check box is checked, dependent variables for the molar surface concentration of the dissolving-depositing species are added. These can be used to model the thickness of an dissolving/depositing layer in a time-dependent simulation where the resulting deformation in the model geometry is small and will have negligible impact on the current distribution. In stationary studies, the molar surface concentration variables will be defined but not solved for by default.
 
    The Density of sites parameter will be used to set up the local surface mass balance equations for the fractional surface coverages.
 
    Use the Add (

) and 
Delete (

) buttons as needed in the table to control the number of species. The name of each adsorbing-desorbing 
Species may be edited in the first column of the table. Use a 
Site occupancy number larger than unity for species that occupy more than one site at the electrode surface.
 
    According to the settings of the table, the fractional surface coverage variable names will be defined according to xxx.theta_yyy_zzz where 
xxx is the interface name, 
yyy the tag of the electrode surface node, and 
zzz the species name. The variable 
xxx.thetafree_yyy represents the fraction of free surface sites. These variable names may be used when defining kinetics parameters in 
Electrode Reaction subnodes.
 
    Specify either a Surface resistance Rfilm (SI unit: 
Ω·m
2) directly or choose the 
Thickness and conductivity option to calculate the surface resistivity based on a depositing film thickness.
 
    The perturbation parameter is either Electric potential, 
Electrode potential, 
Total current, or 
Average current density, based on the 
Boundary condition selected in the next section.
 
    Use the Electric potential option to set the value of the potential explicitly with respect to ground whereas the 
Electrode potential will set the potential value with respect to a reference potential. 
Total current, 
Average current density, and 
External short all add an extra global degree of freedom for the potential in the electrode phase, set to comply with the chosen condition.
 
    When using the Total current option in 1D or 2D, the boundary area is based either on the 
Cross-sectional area (1D), or the 
Out-of-Plane thickness (2D) properties, set on the physics interface top node.
 
    See also the documentation for the Electrode Potential and 
External Short nodes for further information about these boundary condition.
 
    The Cyclic voltammetry setting varies the electric potential linearly in time as follows when used in conjunction with a Cyclic Voltammetry study step:

 
    More advanced waveforms can be obtained using the Electric potential option with a parameter setting based on 
Functions found in the 
Definitions menu.
 
    This boundary condition is only available for the Electroanalysis charge conservation model in the 
Tertiary Current Distribution, Nernst-Planck (tcd) interface.
 
    The Counter electrode option will set a potential to ensure an overall charge balance of the cell so that the integral of all electrode reaction currents of all electrode surface node sums up to zero.
 
    If Cyclic voltammetry is selected as the 
Boundary condition, the 
Smoothing of cyclic voltammetry wave functions check box is selected by default and the 
Smoothing factor defaults to 1·10
-3. When enabled, smoothing is applied on the triangular wave around the vertex potentials. The smoothing zone corresponds to the product of the smoothing factor with half the duration of one period of the triangular wave.
 
    
    This section is only available in the Primary Current Distribution and Secondary Current Distribution interfaces when the Current Distribution Model property has been set to 
Primary. To display this section, click the 
Show More Options button (

) and select 
Advanced Physics Options in the 
Show More Options dialog box.